{"id":11798,"date":"2026-08-25T01:44:38","date_gmt":"2026-08-25T01:44:38","guid":{"rendered":"https:\/\/instrumedpanama.com\/?post_type=product&#038;p=11798"},"modified":"2026-08-25T01:48:24","modified_gmt":"2026-08-25T01:48:24","slug":"newlab-300-pour-point","status":"publish","type":"product","link":"https:\/\/instrumedpanama.com\/en\/product\/newlab-300-pour-point\/","title":{"rendered":"NewLab 300 Pour Point"},"content":{"rendered":"<h5>Measuring Pour Point Devices<\/h5>\n<ul>\n<li>Two PT100 detection probes placed on the surface of the product<\/li>\n<li>Mechanical moving arm bringing the test jar in horizontal position<\/li>\n<\/ul>\n<hr \/>\n<h5>Measuring Temperature Probe<\/h5>\n<ul>\n<li>Platinum resistance PT100 class A<\/li>\n<\/ul>\n<hr \/>\n<h5>Measuring Parameters<\/h5>\n<ul>\n<li>Temperatures: in \u00b0C<\/li>\n<li>Measuring range: -110\u00b0C \u2026 +100\u00b0C<\/li>\n<li>Range of analysis: -90\u00b0C\u2026+60\u00b0C (300\/2-SA)<\/li>\n<li>Resolution: 0.06 \u00b0C<\/li>\n<li>Accuracy: \u00b1 0.1 \u00b0C<\/li>\n<li>Repeatability \/ Reproducibility: as per standards methods or better<\/li>\n<\/ul>\n<hr \/>\n<h5>Software Features<\/h5>\n<p>New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)<\/p>\n<ul>\n<li>User friendly interface<\/li>\n<li>All analytical parameters recorded<\/li>\n<li>Customizable analysis parameters and methods<\/li>\n<li>Customizable results report<\/li>\n<li>Printable graphs and results<\/li>\n<li>Self-identification of the typology of the analysers connected<\/li>\n<\/ul>\n<h5>The software includes:<\/h5>\n<h4>Analysis Menu<\/h4>\n<ul>\n<li>Standard method as per ASTM \/ IP \/ ISO \/ EN \/ DIN\u2026 norms of reference:\n<ul>\n<li>(internal) with sample pre-heating<\/li>\n<li>(external) without sample pre-heating<\/li>\n<\/ul>\n<\/li>\n<li>Optional methods:\n<ul>\n<li>T-sample \u2013 T-bath (Delta T constant)<\/li>\n<li>cooling rate \u00b0C \/ h<\/li>\n<li>selectable bath steps<\/li>\n<li>fast bath<\/li>\n<li>selectable tilt out test temperature<\/li>\n<\/ul>\n<\/li>\n<li>Audible alarm and displayed messages (at the end of the analysis and in case of errors and\/or malfunctions)<\/li>\n<\/ul>\n<h6>Diagnostic Menu<\/h6>\n<ul>\n<li>Direct access to all analog, digital, inputs and outputs<\/li>\n<li>Selectable value displaying: \u00b0C \/ Volt<\/li>\n<\/ul>\n<h6>Calibration Menu<\/h6>\n<ul>\n<li>Automatic calibration of each temperature probe<\/li>\n<li>Last calibration date referred to each single probe displayed and relative data printable<\/li>\n<li>Display of calibration diagram<\/li>\n<li>Insertion of offset values<\/li>\n<li>Standard and advanced calibration modes<\/li>\n<\/ul>\n<h6>Data Utilities<\/h6>\n<ul>\n<li>Fields for introduction of operator and product name<\/li>\n<li>Archive viewer for files recall<\/li>\n<li>All analysis stored in Excel\u00ae compatible format<\/li>\n<li>Storage capacity for more than 60\u2019000 analysis<\/li>\n<li>LIMS compatible<\/li>\n<\/ul>\n<hr \/>\n<h5>Integrated Touch Screen Panel PC<\/h5>\n<ul>\n<li>TFT\/LCD 12&#8243;<\/li>\n<li>Resolution 1024 \u00d7 768, 16.2 M colours<\/li>\n<li>2 USB ports for connection to an external printer and\/or external PC<\/li>\n<li>Storage capacity for more than 60&#8217;000 analysis<\/li>\n<\/ul>\n<hr \/>\n<h5>Test Jar<\/h5>\n<ul>\n<li>Same dimensions and volume as described by the standard test methods<\/li>\n<li>Product level mark<\/li>\n<li>Small edge on the top in order to fix the glass cell to the analytical head<\/li>\n<\/ul>\n<hr \/>\n<h5>Cooling System<\/h5>\n<p>Integrated gas CFC free motor compressors:<\/p>\n<ul>\n<li>Single stage (for temperatures up to -40\u00b0C \/ 1)<\/li>\n<li>Double stage (for temperatures up to -80\u00b0C \/ 2)<\/li>\n<\/ul>\n<p>Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.<\/p>\n<hr \/>\n<h5>Safety Devices<\/h5>\n<ul>\n<li>Pressure controller for 1st stage motor compressor<\/li>\n<li>Pressure controller for 2nd stage motor compressor<\/li>\n<li>Thermostat for 2nd stage activation<\/li>\n<li>Thermo-switch for each cooling \/ heating jacket<\/li>\n<li>Motor compressors equipped with internal overload devices<\/li>\n<\/ul>\n<hr \/>\n<h5>Electrical Supply<\/h5>\n<ul>\n<li>220V \u00b1 15% \/ 50 to 60 Hz<\/li>\n<li>115V \u00b1 15% \/ 60 Hz<\/li>\n<\/ul>\n<hr \/>\n<h5>Cord Cable<\/h5>\n<ul>\n<li>3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives<\/li>\n<\/ul>\n<hr \/>\n<h5>Ambient Temperature<\/h5>\n<ul>\n<li>Max 32 \u00b0C<\/li>\n<li>H.R. 80%<\/li>\n<\/ul>\n<div class=\"table-responsive-sm my-5\">\n<table class=\"table table-sm small table-bordered\">\n<thead class=\"thead-light\">\n<tr>\n<th>Dimensions (cm)<\/th>\n<th>1 test position<\/th>\n<th>2 test position<\/th>\n<th>3 test position<\/th>\n<th>4 test position<\/th>\n<th>6 test position<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>width<\/td>\n<td>66<\/td>\n<td>66<\/td>\n<td>100<\/td>\n<td>134<\/td>\n<td>130<\/td>\n<\/tr>\n<tr>\n<td>depth<\/td>\n<td>60<\/td>\n<td>60<\/td>\n<td>60<\/td>\n<td>60<\/td>\n<td>75<\/td>\n<\/tr>\n<tr>\n<td>height<\/td>\n<td>80<\/td>\n<td>80<\/td>\n<td>80<\/td>\n<td>80<\/td>\n<td>170<\/td>\n<\/tr>\n<\/tbody>\n<thead class=\"thead-light\">\n<tr>\n<th colspan=\"6\">Weight (kg)<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>single stage compressor<\/td>\n<td>70<\/td>\n<td>90<\/td>\n<td>&#8211;<\/td>\n<td>&#8211;<\/td>\n<td>&#8211;<\/td>\n<\/tr>\n<tr>\n<td>double stage compressor<\/td>\n<td>80<\/td>\n<td>100<\/td>\n<td>130<\/td>\n<td>160<\/td>\n<td>280<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<hr \/>\n<h3>NewLab 300 ST<\/h3>\n<ul>\n<li>Measuring range: -110\u00b0C \u2026+100\u00b0C<\/li>\n<li>Range of analysis: -110\u00b0C\u2026+55\u00b0C<\/li>\n<li>Resolution: 0.01 \u00b0C<\/li>\n<li>Width: 34 cm<\/li>\n<li>Depth: 60 cm<\/li>\n<li>Height: 80 cm<\/li>\n<li>Weight: 34 kg<\/li>\n<\/ul>\n<div class=\"row\">\n<div class=\"col-sm-5\"><img data-recalc-dims=\"1\" decoding=\"async\" class=\"img-fluid\" src=\"https:\/\/i0.wp.com\/www.lin-tech.ch\/img\/nl300st.jpg?ssl=1\" alt=\"NewLab 300 ST: Pour Point\" \/><\/div>\n<\/div>","protected":false},"excerpt":{"rendered":"<h6>ASTM D97 &#8211; ASTM D5853 &#8211; ASTM D5950 &#8211; ASTM D6074 &#8211; ASTM D6158<br \/>\nIP 15 &#8211; IP 441<br \/>\nISO 3016<br \/>\nEN ISO 22995<\/h6>\n<p>Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, \u2026<\/p>\n<h5>Measuring Pour Point Principle<\/h5>\n<p>According to the methods, the sample is cooled down at a specified rate and, at the prescribed temperature intervals, the mechanical arm of the analyser lifts the test jar from the cooling jacket and tilts it in order to bring it in horizontal position to test the flow of the product. The sample movement is detected by the thermal probes (PT100 detection) placed above the sample surface which react if touched by the cooled sample.<\/p>","protected":false},"featured_media":11799,"template":"","meta":[],"product_brand":[],"product_cat":[202],"product_tag":[],"class_list":{"0":"post-11798","1":"product","2":"type-product","3":"status-publish","4":"has-post-thumbnail","6":"product_cat-automatic-analysers","8":"first","9":"instock","10":"taxable","11":"shipping-taxable","12":"product-type-simple"},"jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product\/11798","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product"}],"about":[{"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/types\/product"}],"version-history":[{"count":1,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product\/11798\/revisions"}],"predecessor-version":[{"id":11800,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product\/11798\/revisions\/11800"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/media\/11799"}],"wp:attachment":[{"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/media?parent=11798"}],"wp:term":[{"taxonomy":"product_brand","embeddable":true,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product_brand?post=11798"},{"taxonomy":"product_cat","embeddable":true,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product_cat?post=11798"},{"taxonomy":"product_tag","embeddable":true,"href":"https:\/\/instrumedpanama.com\/en\/wp-json\/wp\/v2\/product_tag?post=11798"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}