Hydrogen Sulfide Detection Tubes 2/b (1 – 60 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces.
Hydrogen Sulfide Detection Tubes 2/b (1 – 60 ppm)
Specification:
| Standard Measuring Range: | 2 to 60 ppm |
| Number of strokes n: | 1 |
| Time for Measurement: | approx. 30 sec |
| Standard Deviation: | ± 5 to 10 % |
| Color Change: | white ??> pale brown |
| Temperature: | 0 to 40°C |
| Absolute Humidity: | max. 20 mg H2O / L |
| Reaction Principle: | H2S + Pb2+ ??> PbS + 2 H+ |
| Cross Sensitivity: | Hydrochloric acid, mercaptan and sulfur dioxide in the TLV range does not interfere. |
| Extension of the Measuring Range: |
Using n=2, divide the reading by 2; the measuring range will be 1 to 30 ppm. |