Measuring CFPP Devices
- Aspiration pipette
- Filter assembly
- Light barrier
Measuring Temperature Probe
- Platinum resistance PT100 class A
Internal Vacuum Generator
- 1 x micro-pump of 350 mBar
- 1 x electronic pressure / vacuum regulator composed by: proportional valve, pressure / vacuum control sensor, regulator for reference vacuum generation at 20 mBar, vacuum stabilizer
Measuring Parameters
- Temperatures: in °C
- Measuring range: +80°C … -80°C
- Resolution: 0.06 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
Software Features
New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)
- User friendly interface
- All analytical parameters recorded
- Customizable analysis parameters and methods
- Customizable results report
- Printable graphs and results
- Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
- Optional methods:
- T-sample – T-bath (Delta T constant)
- cooling rate °C / h
- selectable bath steps
- fast bath with selectable temperature
- Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
- The parameters displayed and updated in real time are:
- sample temperature
- bath temperature
- vacuum pressure
- low level light value
- up level light value
- aspiration time
- release time
- intertime test
Thanks to an istogram (graph) that shows the aspiration and release times it is possible to observe the behaviour of the sample during its cooling phase
This feature is an excellent tools for the observation and evaluation of the additivations actions and behaviour
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs
- Selectable value displaying: °C / Volt
- Vacuum data displayed in mBars
Calibration Menu
- Automatic calibration of each temperature probe
- Last calibration date referred to each single probe displayed and relative data printable
- Display of calibration diagram
- Insertion of offset values
- Standard and advanced calibration modes
Data Utilities
- Fields for introduction of operator and product name
- Archive viewer for files recall
- All analysis stored in Excel® compatible format
- Storage capacity for more than 60’000 analysis
- LIMS compatible
Integrated Touch Screen Panel PC
- TFT/LCD 12″
- Resolution 1024 × 768, 16.2 M colours
- 2 USB ports for connection to an external printer and/or external PC
- Storage capacity for more than 60’000 analysis
Cleaning pipette procedure
By using a suitable cleaning liquid and pressing the relevant function button the analyser performs a cleaning sequence of 10 aspirations cycles
Easy removing of aspiration pipette and filter assembly allows cleaning according to the methods
Test Jar
- Same dimensions and volume as described by the standard test methods
- Product level mark
- Small edge on the top in order to fix the glass cell to the analytical head
Cooling System
Integrated gas CFC free motor compressors:
- Single stage (for temperatures up to -40°C / 1)
- Double stage (for temperatures up to -80°C / 2)
Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.
Safety Devices
- Pressure controller for 1st stage motor compressor
- Pressure controller for 2nd stage motor compressor
- Thermostat for 2nd stage activation
- Thermo-switch for each cooling / heating jacket
- Motor compressors equipped with internal overload devices
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives
Ambient Temperature
- Max 32 °C
- H.R. 80%
Spare Parts NewLab 200 SA
Suggested for 2 years operation.
- 1007: Calibrated glass cell for CFPP.
- 1020: Calibrated aspiration pipette for CFPP.
- 3402: Emitter Up (Red).
- 3403: Emitter Down (Blue).
- 3596: PT 100 product w/connector for CFPP.
- 5029: Filter with mesh.
- 6049: Stainless steel filter holder with 10 × mesh interchangeable.
- 7054: O-ring (small) for CFPP filter.
- 7055: O-ring (big) for CFPP filter.
- 7315: O-Ring for Test Jar.
Tools Required for Routine Calibration
Not needed for installation.
- 3013: Calibration decade box – PT100 simulator.
- 3102: Set of connectors and cables.
| Dimensions (cm) | 1 test position | 2 test positions | 3 test positions | 4 test positions | 6 test positions |
|---|---|---|---|---|---|
| width | 66 | 66 | 100 | 134 | 130 |
| depth | 60 | 60 | 60 | 60 | 75 |
| height | 80 | 80 | 80 | 80 | 170 |
| Weight (kg) | |||||
| single stage compressor | 70 | 90 | – | – | – |
| double stage compressor | 80 | 100 | 130 | 160 | 280 |

NewLab 200 ST
- Cooling system ST:
- Integrated gas CFC free Stirling motor compressor
- For temperatures up to -110°C
- Integrated Touch Screen Panel PC for /ST configuration:
- TFT/LCD 8”
- Resolution 1024 × 768
- Measuring range: +55°C … -95°C
- Resolution: 0.01 °C
- Width: 34 cm
- Depth: 60 cm
- Height: 80 cm
- Weight: 34 kg
Spare Parts NewLab 200 ST
Suggested for 2 years operation.
- 1007: Calibrated glass cell for CFPP.
- 1020: Calibrated aspiration pipette for CFPP.
- 3402: Emitter Up (Red).
- 3403: Emitter Down (Blue).
- 3596: PT 100 product w/connector for CFPP.
- 5029: Filter with mesh.
- 6049: Stainless steel filter holder with 10 × mesh interchangeable.
- 7054: O-ring (small) for CFPP filter.
- 7055: O-ring (big) for CFPP filter.
- 7315: O-Ring for Test Jar.
Tools Required for Routine Calibration
Not needed for installation.
- 3013: Calibration decade box – PT100 simulator.
- 3102: Set of connectors and cables.