ASTM D97 - ASTM D5853 - ASTM D5950 - ASTM D6074 - ASTM D6158 IP 15 - IP 441 ISO 3016 EN ISO 22995 Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, …
NewLab 300 Pour Point

NewLab 200 CFPP Cold Filter Plugging Point

ASTM D6371
IP 309
EN 116 – EN 16329

Correlated:

IP 419

Cold Filter Plugging Point of diesel, biodiesel and heating fuels.

Measuring CFPP Principle

The sample is cooled down according to the methods and when the preselected temperature is reached a vacuum of 20 mBar is automatically applied to the sample. The product is sucked through the filter into the calibrated aspiration pipette. If the sample takes more than 60 seconds to reach the upper barrier detector (during the aspiration phase), or it fails to return completely into the test jar before that the product has cooled by a further 1°C, the Cold Filter Plugging Point is reached.

Category: Automatic Analysers
Share:
  • Description
  • Envío y Entrega
Description
Measuring CFPP Devices
  • Aspiration pipette
  • Filter assembly
  • Light barrier
Measuring Temperature Probe
  • Platinum resistance PT100 class A
Internal Vacuum Generator
  • 1 x micro-pump of 350 mBar
  • 1 x electronic pressure / vacuum regulator composed by: proportional valve, pressure / vacuum control sensor, regulator for reference vacuum generation at 20 mBar, vacuum stabilizer
Measuring Parameters
  • Temperatures: in °C
  • Measuring range: +80°C … -80°C
  • Resolution: 0.06 °C
  • Accuracy: ± 0.1 °C
  • Repeatability / Reproducibility: as per standards methods or better

Software Features

New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)

  • User friendly interface
  • All analytical parameters recorded
  • Customizable analysis parameters and methods
  • Customizable results report
  • Printable graphs and results
  • Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
  • Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
  • Optional methods:
    • T-sample – T-bath (Delta T constant)
    • cooling rate °C / h
    • selectable bath steps
    • fast bath with selectable temperature
  • Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
  • The parameters displayed and updated in real time are:
    • sample temperature
    • bath temperature
    • vacuum pressure
    • low level light value
    • up level light value
    • aspiration time
    • release time
    • intertime test

Thanks to an istogram (graph) that shows the aspiration and release times it is possible to observe the behaviour of the sample during its cooling phase

This feature is an excellent tools for the observation and evaluation of the additivations actions and behaviour

Diagnostic Menu
  • Direct access to all analog, digital, inputs and outputs
  • Selectable value displaying: °C / Volt
  • Vacuum data displayed in mBars
Calibration Menu
  • Automatic calibration of each temperature probe
  • Last calibration date referred to each single probe displayed and relative data printable
  • Display of calibration diagram
  • Insertion of offset values
  • Standard and advanced calibration modes
Data Utilities
  • Fields for introduction of operator and product name
  • Archive viewer for files recall
  • All analysis stored in Excel® compatible format
  • Storage capacity for more than 60’000 analysis
  • LIMS compatible
Integrated Touch Screen Panel PC
  • TFT/LCD 12″
  • Resolution 1024 × 768, 16.2 M colours
  • 2 USB ports for connection to an external printer and/or external PC
  • Storage capacity for more than 60’000 analysis

Cleaning pipette procedure

By using a suitable cleaning liquid and pressing the relevant function button the analyser performs a cleaning sequence of 10 aspirations cycles
Easy removing of aspiration pipette and filter assembly allows cleaning according to the methods

Test Jar
  • Same dimensions and volume as described by the standard test methods
  • Product level mark
  • Small edge on the top in order to fix the glass cell to the analytical head
Cooling System

Integrated gas CFC free motor compressors:

  • Single stage (for temperatures up to -40°C / 1)
  • Double stage (for temperatures up to -80°C / 2)

Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.

Safety Devices
  • Pressure controller for 1st stage motor compressor
  • Pressure controller for 2nd stage motor compressor
  • Thermostat for 2nd stage activation
  • Thermo-switch for each cooling / heating jacket
  • Motor compressors equipped with internal overload devices
Electrical Supply
  • 220V ± 15% / 50 to 60 Hz
  • 115V ± 15% / 60 Hz
Cord Cable
  • 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives
Ambient Temperature
  • Max 32 °C
  • H.R. 80%
Spare Parts NewLab 200 SA

Suggested for 2 years operation.

  • 1007: Calibrated glass cell for CFPP.
  • 1020: Calibrated aspiration pipette for CFPP.
  • 3402: Emitter Up (Red).
  • 3403: Emitter Down (Blue).
  • 3596: PT 100 product w/connector for CFPP.
  • 5029: Filter with mesh.
  • 6049: Stainless steel filter holder with 10 × mesh interchangeable.
  • 7054: O-ring (small) for CFPP filter.
  • 7055: O-ring (big) for CFPP filter.
  • 7315: O-Ring for Test Jar.
Tools Required for Routine Calibration

Not needed for installation.

  • 3013: Calibration decade box – PT100 simulator.
  • 3102: Set of connectors and cables.
Dimensions (cm) 1 test position 2 test positions 3 test positions 4 test positions 6 test positions
width 66 66 100 134 130
depth 60 60 60 60 75
height 80 80 80 80 170
Weight (kg)
single stage compressor 70 90 – – –
double stage compressor 80 100 130 160 280

NewLab 200 ST: Cold Filter Plugging Point - CFPP

NewLab 200 ST

  • Cooling system ST:
    • Integrated gas CFC free Stirling motor compressor
    • For temperatures up to -110°C
  • Integrated Touch Screen Panel PC for /ST configuration:
    • TFT/LCD 8”
    • Resolution 1024 × 768
  • Measuring range: +55°C … -95°C
  • Resolution: 0.01 °C
  • Width: 34 cm
  • Depth: 60 cm
  • Height: 80 cm
  • Weight: 34 kg

Spare Parts NewLab 200 ST

Suggested for 2 years operation.

  • 1007: Calibrated glass cell for CFPP.
  • 1020: Calibrated aspiration pipette for CFPP.
  • 3402: Emitter Up (Red).
  • 3403: Emitter Down (Blue).
  • 3596: PT 100 product w/connector for CFPP.
  • 5029: Filter with mesh.
  • 6049: Stainless steel filter holder with 10 × mesh interchangeable.
  • 7054: O-ring (small) for CFPP filter.
  • 7055: O-ring (big) for CFPP filter.
  • 7315: O-Ring for Test Jar.
Tools Required for Routine Calibration

Not needed for installation.

  • 3013: Calibration decade box – PT100 simulator.
  • 3102: Set of connectors and cables.
Envío y Entrega
wd-ship-2

ENVÍO Y ENTREGA

Tenemos un enfoque centrado en el cliente y priorizamos los procesos de logística y transporte —y todos los necesarios— para aumentar la satisfacción de las empresas que compran nuestros productos.

Related products

Configuration Table

Automatic Analysers
Configuration Table PP CP CFPP FP CP+PP
Read more
NewLab 100 Cloud Point

NewLab 100 Cloud Point

Automatic Analysers
ASTM D5771 DIN 51597 EN 23015 EN 590 IP 444 Correlated: ASTM D2500 - ASTM D5772 - ASTM D5773 IP 219 - IP 445 - IP 446 ISO 3015 JIS K2269
Cloud Point of petroleum products and biodiesel fuels
Read more
ASTM D97 - ASTM D5853 - ASTM D5950 - ASTM D6074 - ASTM D6158 IP 15 - IP 441 ISO 3016 EN ISO 22995 Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, …

NewLab 300 Pour Point

Automatic Analysers
ASTM D97 - ASTM D5853 - ASTM D5950 - ASTM D6074 - ASTM D6158 IP 15 - IP 441 ISO 3016 EN ISO 22995
Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, …
Measuring Pour Point Principle
According to the methods, the sample is cooled down at a specified rate and, at the prescribed temperature intervals, the mechanical arm of the analyser lifts the test jar from the cooling jacket and tilts it in order to bring it in horizontal position to test the flow of the product. The sample movement is detected by the thermal probes (PT100 detection) placed above the sample surface which react if touched by the cooled sample.
Read more
instrumed_white
  • Inicio
  • Nosotros
  • Productos
  • Contacto
INSTRUMED PANAMA 2023 CREADO POR PERFILES LATAM.
payments
  • Inicio
  • Productos
    • HORIBA
    • ThermoProbe
    • OMNITEK
  • Nosotros
  • Contacto
  • Compare
  • Login / Register
Shopping cart
close

Social Chat is free, download and try it now here!

English
Spanish