Measuring Pour Point Devices
- Two PT100 detection probes placed on the surface of the product
- Mechanical moving arm bringing the test jar in horizontal position
Measuring Temperature Probe
- Platinum resistance PT100 class A
Measuring Parameters
- Temperatures: in °C
- Measuring range: -110°C … +100°C
- Range of analysis: -90°C…+60°C (300/2-SA)
- Resolution: 0.06 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
Software Features
New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)
- User friendly interface
- All analytical parameters recorded
- Customizable analysis parameters and methods
- Customizable results report
- Printable graphs and results
- Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference:
- (internal) with sample pre-heating
- (external) without sample pre-heating
- Optional methods:
- T-sample – T-bath (Delta T constant)
- cooling rate °C / h
- selectable bath steps
- fast bath
- selectable tilt out test temperature
- Audible alarm and displayed messages (at the end of the analysis and in case of errors and/or malfunctions)
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs
- Selectable value displaying: °C / Volt
Calibration Menu
- Automatic calibration of each temperature probe
- Last calibration date referred to each single probe displayed and relative data printable
- Display of calibration diagram
- Insertion of offset values
- Standard and advanced calibration modes
Data Utilities
- Fields for introduction of operator and product name
- Archive viewer for files recall
- All analysis stored in Excel® compatible format
- Storage capacity for more than 60’000 analysis
- LIMS compatible
Integrated Touch Screen Panel PC
- TFT/LCD 12″
- Resolution 1024 × 768, 16.2 M colours
- 2 USB ports for connection to an external printer and/or external PC
- Storage capacity for more than 60’000 analysis
Test Jar
- Same dimensions and volume as described by the standard test methods
- Product level mark
- Small edge on the top in order to fix the glass cell to the analytical head
Cooling System
Integrated gas CFC free motor compressors:
- Single stage (for temperatures up to -40°C / 1)
- Double stage (for temperatures up to -80°C / 2)
Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.
Safety Devices
- Pressure controller for 1st stage motor compressor
- Pressure controller for 2nd stage motor compressor
- Thermostat for 2nd stage activation
- Thermo-switch for each cooling / heating jacket
- Motor compressors equipped with internal overload devices
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives
Ambient Temperature
- Max 32 °C
- H.R. 80%
| Dimensions (cm) | 1 test position | 2 test position | 3 test position | 4 test position | 6 test position |
|---|---|---|---|---|---|
| width | 66 | 66 | 100 | 134 | 130 |
| depth | 60 | 60 | 60 | 60 | 75 |
| height | 80 | 80 | 80 | 80 | 170 |
| Weight (kg) | |||||
| single stage compressor | 70 | 90 | – | – | – |
| double stage compressor | 80 | 100 | 130 | 160 | 280 |
NewLab 300 ST
- Measuring range: -110°C …+100°C
- Range of analysis: -110°C…+55°C
- Resolution: 0.01 °C
- Width: 34 cm
- Depth: 60 cm
- Height: 80 cm
- Weight: 34 kg