NewLab 100 Cloud Point

ASTM D5771 DIN 51597 EN 23015 EN 590 IP 444 Correlated: ASTM D2500 - ASTM D5772 - ASTM D5773 IP 219 - IP 445 - IP 446 ISO 3015 JIS K2269
Cloud Point of petroleum products and biodiesel fuels

NewLab 300 Pour Point

ASTM D97 - ASTM D5853 - ASTM D5950 - ASTM D6074 - ASTM D6158 IP 15 - IP 441 ISO 3016 EN ISO 22995
Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, …
Measuring Pour Point Principle
According to the methods, the sample is cooled down at a specified rate and, at the prescribed temperature intervals, the mechanical arm of the analyser lifts the test jar from the cooling jacket and tilts it in order to bring it in horizontal position to test the flow of the product. The sample movement is detected by the thermal probes (PT100 detection) placed above the sample surface which react if touched by the cooled sample.

NewLab 200 CFPP Cold Filter Plugging Point

ASTM D6371 IP 309 EN 116 - EN 16329

Correlated:

IP 419
Cold Filter Plugging Point of diesel, biodiesel and heating fuels.
Measuring CFPP Principle
The sample is cooled down according to the methods and when the preselected temperature is reached a vacuum of 20 mBar is automatically applied to the sample. The product is sucked through the filter into the calibrated aspiration pipette. If the sample takes more than 60 seconds to reach the upper barrier detector (during the aspiration phase), or it fails to return completely into the test jar before that the product has cooled by a further 1°C, the Cold Filter Plugging Point is reached.